|
|
| Model Number |
N/A |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Occupation: lawyer.Pma/510(k) #: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was reported by a law firm "in 2016, (patient) suffered from varicose veins of the lower extremities and pelvic congestion syndrome." patient elected to have an embolization procedure, during which several cook nester and tornado embolization coils were implanted.Report states the patient "experienced a litany of health issues until she had the coils removed, including: extreme pain; autoimmune diagnoses; difficulty walking; and renal issues.On or about (b)(6) 2018, patient opted to have the coils removed, at which time it was found the coils had perforated and migrated.Report further states "after the coils were removed, patient began to feel relief." medical records indicated the following: on (b)(6) 2016, patient underwent bilateral coiling of gonadal veins due to reflux and dilation.Bilateral ovarian vein insufficiency with markedly enlarged ovarian veins and pelvic varices was confirmed.Successful bilateral ovarian vein coil embolization distal to proximal was performed and periodic venograms confirmed stasis and closure of ovarian vein branches.There were no immediate complications and upon removal of the catheter and sheath, hemostasis was achieved with manual pressure.Patient reportedly experienced some time of relief but the pain came back and became more progressive.A repeat venogram was performed which did not find a significant gradient across left renal vein and no intervention was done at that time.Imaging found no flow to her gonadal veins and no significant compression of her left renal vein.However, patient continued to experience progressive chronic left flank and groin pain and less severe pain on the right side.On (b)(6) 2018, patient underwent marcaine instillation into the ureter to provide data as to whether an auto-transplant of her kidney could potentially cure her pain.The procedure was completed without complications and once awake in the pacu, the patient experienced complete resolution of her left flank pain.The return of pain coincided with the half-life of marcaine.On (b)(6) 2018, patient presented for nephrectomy with auto transplantation as well as removal of her bilateral coils "which have failed to provide her any relief." during the procedure, the laparoscopic left nephroureterectomy was "at least 50% more difficult than the average laparoscopic nephroureterectomy due to her previous gonadal vein coilings performed." there was "chronic scarring and inflammatory tissue in the renal hilum , retroperitoneum and pelvis" that required at least 2 hours longer operative time to complete this portion of the operation successfully and without complication.A bilateral gonadal vein resection and foreign body removal was conducted successfully along with kidney autotransplant, ureterotomoy with ureteral implant and open ureteral stent, bilateral ovary pexy to pelvic side wall, and excision of uterine mass.The entire left gonadal vein and its coils, and the distal right gonadal vein and coils were sent to pathology.On (b)(6) 2018 surgical pathology results notes examination of "the left gonadal vein with foreign body" it is listed as a 6.5x1.5x1cm vein.The surface was described as tan-pink and smooth with metal coils visible from within the lumen.The vein was opened to reveal that the vessel is filled with metallic coiled object measuring 4.5x1x1 cm.The metal coil was noted as silver in color.An examination of the "right gonadal vein with foreign body" was also described.It is listed as a 7.5x1.5x1cm piece of venous tissue with tan-pink surface, and transparent from the lumen to the surface is a coiled metallic silver-colored fragment that measures 7x1x0.5cm.On (b)(6) 2018 postoperative follow up by physician notes patient doing very well status post nephro ureterectomy, foreign body removal of the gonadal veins, uterine mass removal and renal autotransplant.Her presenting chronic flank and back pain was resolved.Appropriate surgical pain was noted.No other adverse effects were reported for this incident.The following are reports that capture different coils implanted into the patient in this event: medwatch report with patient identifier (b)(4) lot #4104782, quantity 1, medwatch report with patient identifier (b)(4) lot# 5849807, quantity 1, medwatch report with patient identifier (b)(4) lot #6431726, quantity 2, medwatch report with patient identifier (b)(4) lot #5769806, quantity 1, medwatch report with patient identifier (b)(4) lot # 6747110, quantity 2, medwatch report with patient identifier (b)(4) lot # 6384575, quantity 6, medwatch report with patient identifier (b)(4) lot # 5793360, quantity 1, medwatch report with patient identifier (b)(4) lot # 6731968, quantity 1, medwatch report with patient identifier (b)(4) lot#6533132, quantity 1, medwatch report with patient identifier (b)(4) lot#3649060, quantity 1, medwatch report with patient identifier (b)(4) lot#3655472, quantity 1, medwatch report with patient identifier (b)(4) lot#4544765, quantity 1 (this report).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ¿ evaluation: a law firm reported an incident involving several nester and tornado embolization coils from different lots.The 44 year-old female patient presented with varicose veins of the lower extremities and pelvic congestion, and had been diagnosed with may-thurner and ehlers-danlos syndromes.She elected to have bilateral coiling of gonadal veins on (b)(6) 2016 due to reflux and dilation.Multiple nester and tornado embolization coils were successfully placed in the ovarian veins.The patient reportedly experienced a period of relief, but began to experience various medical issues.A physician suggested the coils could be contributing to the issues, and suggested they be removed for metal intolerance and toxicity.On (b)(6) 2019, the patient underwent bilateral gonadal vein resection and coil removal, kidney autotransplant, ureterotomoy with ureteral implant and open ureteral stent, bilateral ovary pexy to pelvic side wall, and excision of a uterine mass.The coils had allegedly perforated and migrated.The patient felt relief after removal of the coils.The patient is further stating that the coils are labeled to be 100% platinum, but contain trace amounts of tungsten.Cook received the patient's medical records, implant and explant records, and pathology report on (b)(6) 2019.The implant record listed several lots of cook embolization coils.(b)(6) 2021, additional lot numbers were provided.This investigation addresses lot 4544765, one of the eleven nester lots reported.A review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device, as well as imaging review, were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, the additional information provided on 10dec2019 included x-ray imaging of the patient¿s abdomen that was provided in the explant report.The image shows embolization coils in the patient¿s pelvic veins.This was submitted for medical imaging review.The imaging review concluded that ¿bilateral embolization coils in the expected distribution of the ovarian veins were normal.¿ additional discussion with medical affairs and product management concluded that there is no indication that the coils perforated and migrated, as they remained in the intended location of the pelvic vein as displayed in the imaging.The slight uncoiling of the proximal coil in the left vein could have occurred during coil placement and is not unexpected according to product management.Furthermore, the pathology report of the resected gonadal veins indicates that the embolization coils remained in the target location.At this time, there is no evidence that the complaint device is nonconforming.Additionally, a document based investigation evaluation was performed.The risks associated with these devices are acceptable when weighed against the benefits.Proper inspection activities are in place to identify nonconforming product prior to lot release.A review of the design history file showed that biocompatibility testing of nester coils has been completed as described in iso standards.The device history record of the complaint device lot was reviewed.The dhr for this lot recorded no related non-conformances.A database search revealed no other complaints have been reported for the device lot.As there are adequate inspection activities established, objective evidence that the dhr was fully executed, no non-conformances and no additional complaints from the same lot, it was concluded that there is no evidence that nonconforming product exists in house or in field.The product labeling was reviewed.The product ifu provides the following information to the user related to the reported failure mode: device description: "nester embolization coils are made of platinum with spaced synthetic fibers, and are supplied preloaded in a loading cartridge.They are designed to be delivered to the target vessel using a soft, straight wire guide through a standard angiographic catheter." warnings: "positioning of the embolization coils should be done with particular care.Coils should not be left too close to the inlets of arteries and should be intermeshed with previously placed coils if possible.A minimal but sufficient arterial blood flow should remain to hold the coils against the previously placed coils until a solid clot ensures permanent fixation.The purpose of these suggestions is to minimize the possibility of loose coils becoming dislodged and obstruction a normal and essential arterial channel." instructions for use: "perform a final angiogram to confirm the coil position within the target vessel." how supplied: "supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred." the patient presented with may-thurner (nutcracker) syndrome, pelvic congestion syndrome, ehlers-danlos syndrome, pelvic varices, chronic retroperitoneal and flank venous hypertension, loin pain hematuria syndrome, and uterine fibroid mass in addition to having previously placed bilateral cook nester coils.The patient underwent a bilateral gonadal vein resection and coil removal, kidney auto-transplant, ureterectomy with ureteral implant and open ureteral stent, bilateral ovary pexy to pelvic side wall, and excision of uterine mass in a single procedure.During the procedure, the surgeon noted inflammation in the gonadal veins, adhesions in the renal hilum, and long-standing venous outflow obstruction due to engorged variceal-type perinephric veins draining into the perinephric fat.The patient reportedly experienced instant relief from her symptoms after this procedure.The patient had various pre-existing conditions which could have caused or contributed to chronic pelvic pain.Therefore, cook is unable to attribute the patient's symptoms to any specific cause.Based on the information provided, no returned product, and the results of the investigation, a definitive cause could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|
Search Alerts/Recalls
|
|
|