• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1014261-200
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/31/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.Balloon material ruptures can be affected by numerous factors such as balloon damage during processing of the balloon material, materials, inflation technique, inflation over rated burst pressure, interactions with other devices, anatomical conditions, a previously implanted stent or insufficient preparation prior to use.The investigation was unable to determine a conclusive cause for the reported balloon rupture.It is possible that during advancement the balloon became compromised and/or damaged against the mildly calcified de novo lesion in the anterior tibial artery and/or other devices used in the procedure resulting in the reported balloon rupture; however, this could not be confirmed.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the procedure was performed to treat a mildly calcified de novo lesion in the anterior tibial artery.A 3.0x200mm armada 18 balloon catheter was prepared per the instructions for use; however, the balloon ruptured during the first inflation at 13 atmospheres.The procedure was successfully completed with a non-abbott balloon catheter.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11187772
MDR Text Key228027035
Report Number2024168-2021-00539
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K151317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number1014261-200
Device Lot Number0090241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-