H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one tri-funnel replacement gastrostomy tube 20f was returned for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The investigation is confirmed for the reported balloon rupture and the identified fluid leak issues, as a compound split was noted approximately 1.7 cm from the distal end of the device.The edges of the compound split noted approximately 1.7 cm from the distal end of the device appeared smooth.The balloon was inflated with approximately 10 ml of water and immediately leaked through the rupture.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|