• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1716000J
Device Problems Material Discolored (1170); Fluid/Blood Leak (1250); Fracture (1260); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.The catalog number identified in section has not been cleared in the us but is similar to the powerport slim implantable port, chronoflex single-lumen, 6f products that are cleared in the us.The pro code and 510k number for the powerport slim implantable port, chronoflex single-lumen, 6f products are identified.
 
Event Description
It was reported that post port device placement and during a chemotherapy procedure, the device allegedly had a leak.It was further reported that the physician suspected an occlusion of the catheter.There was no reported patient injury.
 
Event Description
It was reported that some time post port placement through the left subclavian vein, the device allegedly leaked.The distal segment of catheter and the port system were removed.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: one titanium low profile powerport attached to a catheter was returned for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The investigation is confirmed for the reported fluid leak and the identified split and discoloration issue, as a longitudinal split was noted approximately 6.8 cm from the distal end of the cath-lock, discoloration noted around split area and upon infusion of the port body with attached catheter segment, a leak from the longitudinal split was observed.The characteristics of the break are indicative of pinch-off.Pinch-off is known to occur when implanting the catheter into the subclavian vein medial to the border of the first rib, which the user is instructed to avoid in the instructions for use.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Instructions for use states: the risk of pinch-off syndrome can be avoided by inserting the catheter via the internal jugular vein (ij).Subclavian insertion of the catheter medial to the border of the first rib may cause catheter pinch-off, which in turn results in occlusion causing port system failure during power injection.If we choose to insert the catheter into the subclavian vein, it should be inserted lateral to the border of the first rib or at the junction with the axillary vein because such insertion will avoid compression of the catheter, which can cause damage and even sever the catheter.The use of image guidance upon insertion is strongly recommended.A radiographic confirmation of catheter insertion should be made to ensure that the catheter is not being pinched.H10: g3 h11: h6(device, method, result and conclusion) h10: the catalog number identified in section d4 has not been cleared in the us but is similar to the power port slim implantable port, chronoflex single-lumen, 6f products that are cleared in the us.The pro code and 510 k number for the powerport slim implantable port, chronoflex single-lumen, 6f products are identified in d2 and g4.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11192130
MDR Text Key227734234
Report Number3006260740-2020-21382
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
PMA/PMN Number
K072549
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1716000J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Initial Date Manufacturer Received 12/23/2020
Initial Date FDA Received01/19/2021
Supplement Dates Manufacturer Received06/29/2021
Supplement Dates FDA Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-