• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED IMPLANT FOR MANDIBLE, LARGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED IMPLANT FOR MANDIBLE, LARGE Back to Search Results
Model Number SD980.109
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 12/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation ongoing.
 
Event or Problem Description
Mandible implant was wearing through the patients soft tissue and was explanted.
 
Additional Manufacturer Narrative
Suboptimal patient condition after previous surgery with standard implant and bad soft tissue condition of the patient are the root causes of the implant wearing through the patients soft tissue.H3: device was discarded at hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH
Common Device Name
TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED IMPLANT FOR MANDIBLE, LARGE
Manufacturer (Section D)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE  3001
MDR Report Key11192717
Report Number3003998208-2021-00003
Device Sequence Number14387672
Product Code JEY
UDI-Device Identifier05420060381096
UDI-Public05420060381096
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K173039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSD980.109
Device Catalogue NumberSD980.109
Device Lot NumberMU20OXATEC
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2020
Supplement Date Received by Manufacturer02/22/2021
Initial Report FDA Received Date01/19/2021
Supplement Report FDA Received Date03/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age51 YR
-
-