| Model Number |
2067L |
| Medical Device Problem Codes |
Corroded (1131); Moisture Damage (1405)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: it was determined at analysis that the radiofrequency (rf) head failed to recognize devices and failed uplink tests.It passed functional and bench testing using a known good cable.The rf head had internal corrosion and was scrapped.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the radiofrequency programmer head originally returned for service with no information and subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
|
| |
|