|
(b)(4).Batch #: t5ed7w.(b)(4).Device analysis: the product was returned and visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tlc55 device was received with no apparent damage and with a reload loaded in the device.The reload was returned fully fired.The device was tested for functionality with a test reload and it fired, cut, and formed all the staples as intended.The staple line and cut line were complete, and the staples meet the staple form release criteria.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: when positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc., are within the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.The event described could not be confirmed as the device performed without any difficulties.As part of ethicon endo surgerys quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|