• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® PROPATEN® VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® PROPATEN® VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HT087080A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); No Code Available (3191)
Date of Event 12/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2020, the patient presented with an unknown etiology in the femoral artery and underwent treatment utilizing a gore® propaten® vascular graft.The physician completed the implanting procedure successfully.While the patient was in post operative recovery, it was observed that the gore® propaten® vascular graft had reportedly torn above the distal anastomosis post operatively.Further information has been requested but is not available at this time.
 
Additional Manufacturer Narrative
The following fields were updated to reflect additional information received: h6: clinical code, impact codes, type of investigation, investigation findings and investigation conclusions updated to reflect additional information received and completion of investigation.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2020, the patient presented with peripheral arterial disease in the femoral arteries and underwent an aortobifemoral bypass utilizing a gore® propaten® vascular graft.The physician completed the implanting procedure successfully.While the patient was in post operative recovery, it was observed that the patient was bleeding and was taken back to the operating room.The physician observed that the gore® propaten® vascular graft appeared torn from the anastomosis in the profunda post operatively.The physician made the following observations: part of the gore® propaten® vascular graft was still connected to the artery, and when the gore® propaten® vascular graft reportedly broke it retracted back into the tunnel that was created for placement.The physician then decided to ligate the gore® propaten® vascular graft and closed off the profunda.The procedure was completed by performing an above knee amputation.The patient tolerated the procedure and is reported to be recovering.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® PROPATEN® VASCULAR GRAFT
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WEST B/P
1505 n. fourth street
flagstaff AZ 86004
Manufacturer Contact
liam schultz
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key11194486
Report Number2017233-2021-01608
Device Sequence Number14377795
Product Code DSY
UDI-Device Identifier00733132606849
UDI-Public00733132606849
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K062161
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2024
Device Model NumberHT087080A
Device Catalogue NumberHT087080A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2020
Supplement Date Received by Manufacturer12/18/2020
Initial Report FDA Received Date01/19/2021
Supplement Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient SexMale
-
-