| Model Number |
ORT200 |
| Medical Device Problem Codes |
Unintended System Motion (1430); Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The drift in the table was traced to the trend dual counterbalance valve.The valve was replaced and the table was retested to verify proper function.The valve was not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
Operating room staff noticed that over the weekend and long holiday, the ort200 table had drifted head upwards.There was no patient involved.
|
| |
|