MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL2 LAT COXA VARA SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL2 LAT COXA VARA SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L93712
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994)
Date of Event 01/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for trochanteric bursitis.Event is not serious and is considered mild.Event is definitely not related device and possibly related to procedure.Date of implant: (b)(6) 2019, date of event: (b)(6) 2021, (left hip).
 
Event or Problem Description
Clinic note dated (b)(6) 2021: patient reports pain with compression in the left hip secondary to trochanteric bursitis.Patient denies any pain with walking or any issues with mobility, only reports pain when lying on the operative hip.Physical exam confirms pain on palpation of the hip with excellent mobility and rom.Radiology identifies a stable, well-fixed left hip.Surgeon notes the patient had a previous steroid injection in (b)(6) of 2020, to treat the same issue.The surgeon treated the pain and bursitis with a second steroid injection.There were no complications.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, b7 and h6 (patient).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d3 and h6 (patient).From surgical intervention to serious injury.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A review of provided x-ray images finds nothing indicative of a product problem and do not aid in determining a root cause for the reported problems.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 12
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key11195898
Report Number1818910-2021-01474
Device Sequence Number11867812
Product Code KWA
UDI-Device Identifier10603295168898
UDI-Public10603295168898
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K070554
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3L93712
Device Catalogue Number3L93712
Device Lot Number5317589
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/08/2021
Supplement Date Received by Manufacturer01/25/2021
02/25/2021
Initial Report FDA Received Date01/19/2021
Supplement Report FDA Received Date02/03/2021
02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 32IDX50OD.; ARTICUL/EZE BALL 32 +5 BR.; CORAIL2 LAT COXA VARA SIZE 12.; PINN CAN BONE SCREW 6.5MMX15MM.; PINN CAN BONE SCREW 6.5MMX20MM.; PINNACLE SECTOR II CUP 50MM.; ALTRX NEUT 32IDX50OD; ARTICUL/EZE BALL 32 +5 BR; CORAIL2 LAT COXA VARA SIZE 12; PINN CAN BONE SCREW 6.5MMX15MM; PINN CAN BONE SCREW 6.5MMX20MM; PINNACLE SECTOR II CUP 50MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient Weight57
-
-