• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INTROCAN SAFETY; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INTROCAN SAFETY; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 4251644-02
Medical Device Problem Codes Fail-Safe Problem (2936); Device Handling Problem (3265)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 12/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.It should be noted that although the exact batch number is unknown, the following potential batch numbers were provided: lot 0061735860; manufactured: 10/05/2020, expiration: 09/30/2025.Lot 0061727004; manufactured: 09/23/2020, expiration: 08/31/2025.
 
Event or Problem Description
As reported by the user facility: "a nurse started an iv and it did not work.The safety clip deployed.While cleaning up, the nurse was stuck in the finger by the needle.".
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number 400499104.One used sample without packaging was returned for evaluation.The complaint sample was reviewed by the product smes and extensive damage was noted on the clip.The clip appeared to be crushed to the extent where the arms were forced open.While damage can occur during the processing of the piece, it is unlikely that the defect was caused by the process.All units are 100% visually inspected by the clip inspection station.A piece with that amount of damage would be rejected immediately.Additionally, the needle guard would not fit over the clip if it was damaged in the way that was observed.Therefore, the component would not have passed through the guard load station without triggering a fault.It is the conclusion of the smes that the damage to the clip occurred outside of the b.Braun manufacturing facility.The reported defect was not confirmed.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown 18109
MDR Report Key11196287
Report Number2523676-2021-00012
Device Sequence Number3700979
Product Code FOZ
UDI-Device Identifier04046963166007
UDI-Public04046963166007
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4251644-02
Device Catalogue Number4251644-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Initial Date Received by Manufacturer 01/06/2021
Supplement Date Received by Manufacturer01/06/2021
Initial Report FDA Received Date01/19/2021
Supplement Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-