MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER RIGHT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER RIGHT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 75102240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint: case-(b)(4).
 
Event or Problem Description
It was reported that the locking mechanism between handle and broach is loosening.Fault in device was noticed during use.The procedure was completed with the same device.A delay of 30 min or less was reported.No patient harm was reported.
 
Additional Manufacturer Narrative
H3, h6: it was reported that the locking mechanism between handle and broach is loose.The fault in device was noticed during use.The procedure was completed with the same device.A delay of 30 min or less and no patient harm was reported.The complaint device, used in treatment, was returned for investigation.A visual inspection was performed.The device shows very limited signs of use and no damage is observed on the connector site to the broach.Furthermore, the instrument was tested with a gauge whereby no deviation was detected.One additional complaint was reported for batch a58530.However, this complaint has another reported failure mode as the complained device.A review of the production documentation for the batch in scope did not reveal any deviation from the standard operating procedure.Based on the performed investigations, the reported failure mode could be confirmed.Previous investigations demonstrated that under specific circumstances the double offset adapter 80/45 may disconnect from the rasp during backslapping.An optimized design of the device has been released in order to reduce the occurrence of this issue.This version of the device will be monitored for similar issues.Should additional information become available, this complaint will be reassessed.This investigation is considered closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAA DOUBLE OFFSET ADAPTER RIGHT 80/45
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key11196301
Report Number9613369-2021-00047
Device Sequence Number703575
Product Code JDI
UDI-Device Identifier00885556535394
UDI-Public00885556535394
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75102240
Device Catalogue Number75102240
Device Lot NumberA58530
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2021
Initial Date Received by Manufacturer 12/29/2020
Supplement Date Received by Manufacturer12/29/2020
Initial Report FDA Received Date01/19/2021
Supplement Report FDA Received Date05/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-