• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION CONTEGRA; CONDUIT,VALVED,PULMONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION CONTEGRA; CONDUIT,VALVED,PULMONIC Back to Search Results
Model Number 200H20
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 11/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: the device remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Conduction disturbances are known potential adverse effects associated with any cardiac or thoracic procedure and can be resolved with medical treatment or the implant of a permanent pacemaker (with the risk-benefit ratio in favor of implant of the heart valve).A conduction disturbance does not indicate a device malfunction or potential manufacturing issue.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that immediately post implant of this pulmonary valved conduit, the patient developed complete heart block (chb).Temporary atrioventricular pacing was initiated.Ultimately 8 days post implant, a permanent pacemaker was implanted.No additional adverse patient effects were reported. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTEGRA
Common Device Name
CONDUIT,VALVED,PULMONIC
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key11199122
Report Number2025587-2021-00210
Device Sequence Number14189873
Product Code MWH
UDI-Device Identifier00643169616936
UDI-Public00643169616936
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
H020003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/27/2020
Device Model Number200H20
Device Catalogue Number200H20
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2020
Initial Report FDA Received Date01/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient Weight13
-
-