According to the reporter, during a procedure, the device's rf (radio frequency) sensor fell out.The surgeon noted that during closure of abdominal incision sites that there was a small item on the head butler tray (tray used to hold surgical instruments above the patient head).It was immediately removed from field and identified by surgical tech and rn (registered nurse) as the indicator from the radio frequency surgical sponge.Sponges that had been used were palpated and one was found have a missing indicator.It was pulled from red bucket and bagged with indicator.Wrapper from back table pack with lot number was recovered and items labeled with patient sticker to betaken to materials manager.Nurse manager was informed.Rn spoke to surgical team and all agreed that x-ray was not indicated as all were confident all pieces were accounted for.Wand scan confirmed no retained sponges.There was no patient injury.
|
|
Additional information: g1 (manufacturer name, mfr contact first name, last name, email, phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the gauze was wadded up and very bloody, the green pouch was not clearly seen in the picture, and the rfid tag was laying beside the gauze.It was reported that the device had a component that disengaged.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|