|
Results: the ace68 was kinked approximately 94.0 cm from the hub.The ace68 was stretched approximately 117.0 thru 122.0 cm from the hub.The total length was approximately 133.5 cm.Conclusions: evaluation of the returned ace68 revealed that the distal shaft of the catheter was stretched.This stretch is likely the reported ovalization.If the ace68 is forcefully manipulated against resistance during use, damage such as stretching may occur.During the functional test, the ace68 was connected to a demonstration canister and pump.The pump was powered on and the ace68 was able to aspirate fluid into the canister.Further evaluation of the returned ace68 revealed a kink in the distal shaft.This damage was likely incidental to the reported complaint and may have occurred during packaging of the device for return.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
The patient was undergoing a thrombectomy procedure in the m1 segment of the right middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68), non-penumbra sheath, balloon guide catheter, and penumbra system aspiration pump max 220 (pump max).During the procedure, the physician advanced the ace68 into the target location and turned on the pump max to initiate aspiration; however, it was noticed that the aspiration was low.Therefore, the ace68 was removed and the physician found the ace68 to be ovalized approximately 10 centimeters from the distal tip.The procedure was completed using a new ace68 with the same sheath, guide catheter, and pump max.There was no report of an adverse effect to the patient.
|