No conclusion can be drawn.No evaluation was performed, as the device was discarded.This dm0010faa easydrill cranial perforator with lot number 317/20 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
|
It was reported that the perforator did not stop and cut the dura.It was also reported that a piece of bone lodged into the brain.On follow-up, it was reported that the event happened during a deep brain stimulation procedure and there was a 20-minute delay in the procedure to remove the bone from the brain.It was also reported that the patient need to remain flat for 24 hours.On further follow-up, it was reported that the patient had no further issues post-surgery and the device was discarded.
|