ZIMMER SPINE REDUCTION MONOAXIAL PEDICLE SCREW DIA.5.5 LG 45; INSTINCT JAVA SYSTEM Back to Search Results
Catalog Number 046W0AN65545
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3003853072-2020-00122.
 
Event or Problem Description
It was reported that the threaded shafts of two reduction screws were found bent via x-rays during surgery after being installed.The screws were removed, and are assumed to have been replaced without reported patient impacts.This is report one of two for this event.
 
Additional Manufacturer Narrative
Summary: the complaint is confirmed for two (2) of two (2) returned 5.5x45mm mono rdctn scr in jva (pn 046w0an65545) for the failure of screws bent during insertion.Visual inspection revealed the screw shanks are bent.Medical records were provided in form of x-ray, the provided x-ray shows that the screws shank are bent.Potential cause: the cause of the damage cannot be determined at this time since there is no information available regarding how the screws were being inserted or handled at the time of the damage.Additionally, it is unknown what condition of bone that the patient had as hard bone could also have contributed to this event.Dhr review and related actions: per dhr review, the part was likely conforming when it left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.Device use: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Event or Problem Description
It was reported that the threaded shafts of two reduction screws were found bent via x-rays during surgery after being installed.The screws were removed, and are assumed to have been replaced without reported patient impacts.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCTION MONOAXIAL PEDICLE SCREW DIA.5.5 LG 45
Common Device Name
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
MDR Report Key11202540
Report Number3003853072-2020-00121
Device Sequence Number2270332
Product Code NKB
UDI-Device Identifier00889024341739
UDI-Public(01)00889024341739(10)V160028
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111301
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number046W0AN65545
Device Lot NumberV16002876
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/25/2020
Supplement Date Received by Manufacturer06/09/2021
Initial Report FDA Received Date01/20/2021
Supplement Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight78
-
-