| Model Number |
RLT261414 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
According to the gore® excluder® aaa endoprosthesis featuring c3 deployment system instructions for use, 20mm of healthy proximal aortic seal zone is required for implant of the trunk-ipsilateral leg component.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, this patient underwent endovascular treatment of an endovascular repair of an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis featuring c3 deployment system.The aortic seal zone was only 15mm and did not meet the inclusion criteria of the instructions for use.Therefore, the physician attempted to secure the trunk-ipsilateral leg component to the proximal aortic seal zone with staples.On (b)(6) 2020, the patient presented with the trunk-ipsilateral leg component that had migrated out of proximal aortic neck.The staples that were used in the original procedure did not hold.No aneurysm enlargement has been noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data- g1: manufacturing site and address.
|
| |
|