Catalog Number 72200616SR |
Device Problems
Connection Problem (2900); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case-(b)(4).The sample is under evaluation by the manufacturing site.
|
|
Event Description
|
It was reported that, during set up for surgery, the "mdu powermax elite shaver" was not connecting to the shaver box.The procedure was successfully completed without delay using a back-up device.No patient injuries or other complications were reported.Preliminary results of investigation found a jammed motor which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10 h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found a discolored cable and frayed lanyard.A functional evaluation revealed a blade stall error/jammed motor.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.
|
|
Search Alerts/Recalls
|
|
|