Model Number 1550350-38 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during unpacking of a 3.50x38 xience sierra, it was noted that the stent implant had dislodged (was only partially on the balloon) still remaining on the stent delivery catheter.The device was not used and there was no patient involvement.Another xience sierra was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
A visual and dimensional inspection was performed on the returned device.The reported stent dislodgement was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|