COVIDIEN 12CC SALINE SYRINGE (10CC); SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 8881570121
Medical Device Problem Codes Fluid/Blood Leak (1250); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the prefill syringe is not steady.The syringe is adding air in the dead space and causing potential patient issues.The customer also stated that when the plunger is pushed back, it moves slowly on it's own causing fluid to leak.
 
Additional Manufacturer Narrative
D4: correction originally reported as unknown lot number is now updated to lot # 20f1224.The investigation found that there were no related issues recorded throughout the manufacturing and control processes.The manufacturing records, including autoclave documents, were reviewed for the reported lot and no related event occurred during the overall process for this lot.Also, each lot is released based on an acceptable quality limit (aql) inspection.The product was in conformance to specifications and was released for distribution meeting all established quality assurance acceptance levels.A sample was not returned for evaluation however a video was available for review and the lot number was visible.As seen in the video, when the air is pushed outside the syringe and after letting go of the plunger, the plunger then recedes back into the barrel letting air inside.This confirms the observed phenomenon from the complaint.Following previous similar complaints, a supplier corrective action request (scar) was sent to the supplier of the syringes.The rubber tip is designed with a larger outside diameter than the syringe barrel internal diameter.This design is intentional and prevents any leakage during use while providing a good seal to maintain sterility of the finished product.This design causes the rubber tip to put pressure on the barrel walls.When the syringe goes through the autoclave process, heat is created which likely causes the syringe barrel wall to form over the rubber tip creating this bulge above the 10ml mark where the rubber tip resides.Depending on the actual dimensions of the rubber tip, barrel internal diameter and autoclave processing, the extent of bulge may vary.The bulge is found to be inherent to our sterilization process and therefore, it is not considered a deficiency.The root cause of the phenomenon reported in the complaint is due to our process that are within the aql.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12CC SALINE SYRINGE (10CC)
Common Device Name
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key11205609
Report Number1282497-2021-09858
Device Sequence Number19385677
Product Code NGT
UDI-Device Identifier10884521000261
UDI-Public10884521000261
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8881570121
Device Catalogue Number8881570121
Device Lot Number20F1224
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/04/2021
Supplement Date Received by Manufacturer01/04/2021
Initial Report FDA Received Date01/20/2021
Supplement Report FDA Received Date03/29/2021
Patient Sequence Number1
-
-