| Model Number |
ETLW1624C93E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft was implanted in the endovascular treatment of an abdominal aortic aneurysm greater than 50mm.It was reported that during a follow-up approximately 27 months post index procedure, an endoleak type 1b was noted at the lci.The event was treated with a limb extension to hypogastric on the same day.As per the physician, the cause of the event was anatomy related due to a lack of purchase in lci and tortuously.No additional clinical sequalae were provided and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion; the reported type ib endoleak was confirmed on the film provided.Lack of pre-implant ct¿s did not allow for a thorough assessment of the pre-implant anatomy.Post-implant ct¿s were not available for review and the stent graft in-vivo configuration could not be evaluated.It is possible that the tortuosity in the left common iliac artery was a factor in the occurrence of the endoleak.It is also likely that lack of sufficient seal zone as noted by the physician contributed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|