| Model Number |
00620-CP |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Local Reaction (2035); Nasal Obstruction (2466); Swelling/ Edema (4577)
|
| Date of Event |
12/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The tubing harness is the only component of the plasma harness set that is eto sterilized.Haemonetics obtained device history record for the lot of the reported plasma harness set, sterilization parameters were within specification and within validated parameters.The sample was not available for return to hameonetics for evaluation.Without a physical sample the root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
On december 30, 2020 haemonetics was notified of a donor reaction that fit the description of an ethylene oxide reaction which occurred on the process phase during a procedure, utilizing the pcs®2 plasma collection system and plasma harness set.The reaction occurred during saline infusion, the donor experienced periorbital swelling, nasal congestion and eye watering.
|
| |
|