• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER; VENA CAVA FILTER Back to Search Results
Catalog Number EC500F
Device Problems Migration or Expulsion of Device (1395); Difficult to Remove (1528); Patient Device Interaction Problem (4001)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a lot history review was performed.The sample was not returned to the manufacturer for evaluation; however, medical records were provided for review.The investigation is confirmed for filter perforation of the inferior vena cava and retrieval difficulties; however, the investigation is inconclusive for filter migration.A definitive root cause for the reported event could not be determined.The device is labeled for single use.
 
Event Description
This report summarizes one malfunction.A review of the reported information indicated that model ec500f vena cava filter allegedly experienced migration, difficult to remove and patient device interaction problem.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The (b)(6) male patient was (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECLIPSE FILTER
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key11208470
MDR Text Key228164431
Report Number2020394-2021-80038
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberEC500F
Device Lot NumberGFUI2738
Date Manufacturer Received12/31/2020
Type of Device Usage N
Patient Sequence Number1
-
-