BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 3/120/130; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 357508
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2020
Type of Reportable Event Malfunction
Event or Problem Description
A passeo 18 balloon catheter was chosen for treatment of a moderately calcified lesion (70 percent stenosis degree) in a mildly tortuous ulnar artery.During inflation at nominal pressure, the balloon burst.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed a fine jet of water emerging from the distal part of the balloon when attempting inflation.Microscopic analysis of the balloon surface showed a small pinhole about 6 mm proximal to distal the distal x-ray marker.In close vicinity to the pinhole several scratches were observed.It therefore seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E.Calcified lesion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 3/120/130
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11209195
Report Number1028232-2021-00397
Device Sequence Number3248303
Product Code LIT
UDI-Device Identifier07640119559193
UDI-Public07640119559193
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number357508
Device Catalogue NumberSEE MODEL NO.
Device Lot Number02182140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Initial Date Received by Manufacturer 01/21/2021
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date01/21/2021
Supplement Report FDA Received Date03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-