• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE Back to Search Results
Model Number 720054-02
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient went to the hospital two weeks before surgery, saying the spectra penile prosthesis length was not correct.As a result of the inspection, it was confirmed that the length was not correct.The patient had the spectra penile prosthesis replaced and extended the tip by 1.5 cm.Following the surgery, the patient was stable and improved.The event resolved.
 
Additional Manufacturer Narrative
H3 device evaluation the rear tip extender was not returned for analysis.The spectra cylinders were visually and functionally tested.Both cylinders had sharp instrument/tool damage to the outer tube, which resulted in a hole and exposed metal segments in cylinder body.This damage is consistent with explant damage.Both cylinders passed the bend test with a force readout of less than 1390 grams.The cylinders therefore performed within specification.Product analysis was unable to confirm the reported event.
 
Event or Problem Description
It was reported that the patient went to the hospital two weeks before surgery, saying the spectra penile prosthesis length was not correct.As a result of the inspection, it was confirmed that the length was not correct.The patient had the spectra penile prosthesis replaced and extended the tip by 1.5 cm.Following the surgery, the patient was stable and improved.The event resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Common Device Name
PROSTHESIS PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
MDR Report Key11210426
Report Number2183959-2020-06400
Device Sequence Number14189699
Product Code FAE
UDI-Device Identifier00878953005218
UDI-Public00878953005218
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090663
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/28/2023
Device Model Number720054-02
Device Catalogue Number720054-02
Device Lot Number1000155810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2021
Initial Date Received by Manufacturer 12/29/2020
Supplement Date Received by Manufacturer02/03/2021
Initial Report FDA Received Date01/21/2021
Supplement Report FDA Received Date02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
-
-