| Model Number |
1488010001 |
| Medical Device Problem Code |
Electrostatic Discharge (2149)
|
| Health Effect - Clinical Codes |
Shock from Patient Lead(s) (3162); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the user was shocked.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: nurse was shocked when unplugging camera from ccu.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.Root cause: camera head conforms to specifications.The most provable root cause could be damaged power cord or electrical static discharge when connecting the unit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the user was shocked.
|
| |
|