• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 1488210105
Medical Device Problem Code Electrostatic Discharge (2149)
Health Effect - Clinical Codes Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported the user was shocked.
 
Event or Problem Description
It was reported the user was shocked.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: shock when unplugging.Probable root cause: "isolation circuit failure, power supply, filter / fuse, ac inlet board, manufacturing/ service nonconformity, use error.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11211364
Report Number0002936485-2021-00045
Device Sequence Number3795205
Product Code GCJ
UDI-Device Identifier07613327061055
UDI-Public07613327061055
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K132785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1488210105
Device Catalogue Number1488210105
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/23/2020
Supplement Date Received by Manufacturer12/23/2020
Initial Report FDA Received Date01/21/2021
Supplement Report FDA Received Date04/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-