| Model Number |
ZA9003 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided.Best estimate of date of event is between 10/8/2020 and 12/22/2020.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's left eye (os) due to dysphotopsia.At explant, there was no incision enlargement, no vitrectomy, and no sutures done.The explanted iol was replaced with a non-johnson and johnson iol, lower diopter power (23.5).No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10.Device available for evaluation? yes.Returned to manufacturer on: 1/25/2021.Section h3.Device returned to manufacturer? yes.Device evaluation: visual inspection under magnification revealed, haptic damage and that the lens was received cut in half.Which is consistent with a lens, that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed.And no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed, that the product was manufactured and released according to specifications.A search revealed, no other complaints have been received for this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|