| Model Number |
9735560 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient age not available from the site.The catheter was returned to the manufacture for evaluation.After visual/physical examination the reported issue was confirmed.As reported, the fiber had been broken over and was no longer usable.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used in a soft tissue ablation (neuro).It was reported that the fiber snapped right at the bolt.It had a good return and flow throughout the catheter prior to discovering that the fiber was damaged.There was less than an hour delay in the procedure.No impact on patient outcome.
|
| |
|