• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI LATERAL ENTRY FEMORAL RECON NAIL-EX/400MM/RT-STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI LATERAL ENTRY FEMORAL RECON NAIL-EX/400MM/RT-STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 04.003.360S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/ investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/ or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: manufacturing location: (b)(4), manufacturing date: jul 28, 2009, expiration date: jun- 1, 018, part number: 04.003.360s, 10mm ti lateral entry femoral recon nail-ex/400mm/rt-sterile, lot number: 6182035 (non-sterile).Work order traveler met all inspection acceptance criteria.Inspection sheet, in-process acceptance, (b)(4) met all inspection acceptance criteria.Inspection sheet, inspect dimensional, (b)(4) met all inspection acceptance criteria.Inspection sheet, final inspection, (b)(4) met all inspection acceptance criteria.Packaging label log lppf, lmd/lpf rev d was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.(b)(4) supplied by (b)(4) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal for an unknown reason¿ does not indicate breakage of the nail.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Device history review: this lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent a hardware removal of titanium cannulated lateral entry femoral recon nail, titanium end cap and unknown locking screw due to unknown reason.The procedure was successfully completed.This report is for one (1) 10mm ti lateral entry femoral recon nail-ex/400mm/rt-sterile.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM TI LATERAL ENTRY FEMORAL RECON NAIL-EX/400MM/RT-STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11214627
Report Number2939274-2021-00407
Device Sequence Number2994231
Product Code HSB
UDI-Device Identifier10886982082077
UDI-Public10886982082077
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K040336
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/01/2018
Device Model Number04.003.360S
Device Catalogue Number04.003.360S
Device Lot Number6182035
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/28/2020
Supplement Date Received by Manufacturer01/25/2021
Initial Report FDA Received Date01/22/2021
Supplement Report FDA Received Date02/11/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TI END CAP T40 STARDRIVE 0MM EXT FOR FEMORAL NAILS; UNK - SCREWS: LOCKING; TI END CAP T40 STARDRIVE 0MM EXT FOR FEMORAL NAILS; UNK - SCREWS: LOCKING
Outcome Attributed to Adverse Event Required Intervention;
-
-