• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH MIS SINGLE INNER SETSCW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH MIS SINGLE INNER SETSCW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 186715000
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: additional product codes: kwp; kwq; mnh; mni; osh the device history record (dhr) of product code 186715000, lot 284838, was reviewed and no non-conformances were observed during the manufacturing process.The product was released on september 16, 2020.A product investigation was completed: upon visual inspection, it was observed that threads are torn off.Functional test was performed on the device.Inner set screw was tried to place into screw.An attempt made to assemble both mating devices, it was not placed properly however it got stuck & unable to disassemble since the threads of both the devices set screw and screw were damaged.No dimensional inspection can be performed due to post-manufacturing damage.The relevant drawing was reviewed.There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(6) reports an event as follows: the thread of the screw was turned off by the persuader of the viper system.The inner set screw could not find any thread to lock into.The screw and the inner were both replaced with new devices.The procedure was successfully completed with a ten (10) minute surgical delay.No fragments were generated.During the manufacturer investigation of the returned product it was observed that threads are torn off.This report is for a inner set screw.This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIS SINGLE INNER SETSCW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key11215361
MDR Text Key228221512
Report Number1526439-2021-00174
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034121192
UDI-Public(01)10705034121192
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number186715000
Device Catalogue Number186715000
Device Lot Number284838
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received01/22/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
MIS TI CFX FEN POLY 6X40
-
-