Model Number 9734683
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.The orange instrument tracker was returned to the manufacturer for evaluation.Testing found that the side tabs on the unit had broken off at the tip, preventing retention of an inserted instrument.Functional testing found that the tracker could be tracked.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding navigation and spinal instruments, outside of procedure.It was reported that the orange instrument tracker was damaged.It was reported that the instrument tracker and tap got stuck, it was not possible to remove the tap.No patient was present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11220178
Report Number1723170-2021-00195
Device Sequence Number3343947
Product Code OLO
UDI-Device Identifier00763000076634
UDI-Public00763000076634
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K171267
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734683
Device Catalogue Number9734683
Device Lot Number170417
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2021
Initial Report FDA Received Date01/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-