| Model Number |
02215-220-E |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The driver card pcb card was replaced and the machine was working properly.On (b)(6) 2021, driver pcb sample was returned to haemonetics for manufacturing evaluation.Based on visual inspection the u5 component of the driver pcb was burnt.Manufacturing evaluation determined that the driver pcb failure was caused by u5 component failure, the cause could not be determined.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020 haemonetics was notified of a burnt smell upon start-up on a acp 215 automated cell processing system in singapore.This incident occurred during start-up, there was no donor involved in the incident.
|
| |
|