Model Number 732119
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an unspecified procedure, the reflection acet trial shell handle was for the first time and it did not thread into the trial heads.Instruments outside patient.The threads look bad.Procedure finished with s&n backup device.No surgical delay or injury to patient reported due this event.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The threads on the device are damaged, rendering the device inoperative.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF ACET TRL SHELL HANDLE
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11222188
Report Number1020279-2021-00557
Device Sequence Number17082779
Product Code JDG
UDI-Device Identifier03596010259011
UDI-Public03596010259011
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number732119
Device Catalogue Number732119
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Received by Manufacturer 01/04/2021
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date01/25/2021
Supplement Report FDA Received Date05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No