MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, Back to Search Results
Model Number 71362290
Medical Device Problem Codes Material Deformation (2976); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case: (b)(4).
 
Event or Problem Description
It was reported that, a flexible screwdriver was found to be stripped and worn.Incident occurred during an office inspection.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be stripped and worn, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Further review by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The event states that the flexible screwdriver was found damaged during inspection without patient involvement.Users are instructed to inspect all instruments before use and not to use the device if any damage is found.Device damage that is identified during the inspection process will not cause or contribute to a serious injury or death, nor would it be likely to do so in the event of recurrence as the inspection process itself mitigates the risk of any potential harm occurring.Therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If additional details confirm the occurrence of a reportable event, we will update our records accordingly and submit a subsequent report detailing both the event and our complete investigations.While this event is no longer considered reportable, this report is being submitted as a courtesy to provide updated investigation results based on the return and evaluation of the complaint device subsequent to the submission of our previous supplemental report.The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be stripped and worn, rendering the device inoperable.The device shows signs of extensive use.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed part number, this failure mode will be monitored for future complaints for any necessary corrective actions.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.The device should not be reprocessed or reused if comes in contact with blood, tissue or bodily fluids.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SCREWDRIVER
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11222278
Report Number1020279-2021-00560
Device Sequence Number14391767
Product Code HRY
UDI-Device Identifier03596010454171
UDI-Public03596010454171
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362290
Device Catalogue Number71362290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2020
Supplement Date Received by Manufacturer01/23/2021
03/03/2026
Initial Report FDA Received Date01/25/2021
Supplement Report FDA Received Date01/30/2021
03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-