Model Number 82666
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was due to plasma line occlusion.The centrifuge stop during blood prime due to 'centrifuge pressure too high' alert may have introduced an airblock in the plasma centrifuge line.Plasma line occlusion results in more volume being drawn through the lrs chamber than intended, which disrupts leukoreduction.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6).The platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additiona information in h.6 and h.10 investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause : the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was due to plasma line occlusion.The centrifuge stop during blood prime due to 'centrifuge pressure too high' alert may have introduced an airblock in the plasma centrifuge line.Plasma line occlusion results in more volume being drawn through the lrs chamber than intended, which disrupts leukoreduction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL LRS PLATELET SAMPLER, PLASMA, AUTO RBC
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key11223172
Report Number1722028-2021-00025
Device Sequence Number3712003
Product Code GKT
UDI-Device Identifier05020583826663
UDI-Public05020583826663
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2022
Device Model Number82666
Device Catalogue Number82666
Device Lot Number2012032130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2021
Supplement Date Received by Manufacturer02/11/2021
Initial Report FDA Received Date01/25/2021
Supplement Report FDA Received Date02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-