| Model Number |
1867 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the collar was fractured.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A guidezilla guide extension catheter was selected for use.During procedure, it was noted that the collar was fractured.The device was simply pulled out completely from the patient's body.The procedure was completed with another of the same device.No complications reported and patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.Analysis of the tip, distal shaft, collar and hypotube included microscopic and visual inspection.Inspection revealed a partial separation at the shaft/collar area, and the collar was damaged (edge lifted away from collar).Inspection of the rest of the device found no other damage or defect.
|
| |
|
Event or Problem Description
|
|
It was reported that the collar was fractured.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified right coronary artery.A guidezilla guide extension catheter was selected for use.During procedure, it was noted that the collar was fractured.The device was simply pulled out completely from the patient's body.The procedure was completed with another of the same device.No complications reported and patient was stable.
|
| |
|