BECTON DICKINSON CARIBE LTD. BD BACTEC STANDARD ANAEROBIC/F CULTURE VIALS; SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442191
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported while testing with bd bactec¿ standard anaerobic/f culture vials the bottles are flagged as positive.Gram stain shows negative.There was no reported patient impact.
 
Event or Problem Description
It was reported while testing with bd bactec¿ standard anaerobic/f culture vials the bottles are flagged as positive.Gram stain shows negative.There was no reported patient impact.
 
Additional Manufacturer Narrative
H6: investigation summary: customer reported a false positive defect.Bd was unable to reproduce customer experience with the bactec product.A false positive response was not observed when retention samples were tested.Batch history records were reviewed, and all testing were within specification for product release.Users are cautioned in the package insert under limitation of the procedure: ¿a gram-stained smear from culture medium may contain small number of non-viable organisms derived from media constituents, staining reagents, immersion oil, glass slide and specimens used for inoculation.There are many factors that can influence the false positive rate, including blood volume, blood cell counts, environmental factors, and media lot to lot variations.Complaint is unconfirmed based on retention samples and batch history record review.No correctives actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.H3 other text: see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC STANDARD ANAEROBIC/F CULTURE VIALS
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
MDR Report Key11223711
Report Number3008352382-2021-00006
Device Sequence Number4228312
Product Code MDB
UDI-Device Identifier00382904421915
UDI-Public0382904421915
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K915796
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2021
Device Catalogue Number442191
Device Lot Number0237177
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/05/2021
Supplement Date Received by Manufacturer02/14/2021
Initial Report FDA Received Date01/25/2021
Supplement Report FDA Received Date02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-