MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO REFLECT METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO REFLECT METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Model Number 023-878
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Fatigue (1849); Headache (1880); Hypoglycemia (1912); Shaking/Tremors (2515)
Date of Event 01/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2021, the lay-user/patient contacted lifescan (lfs) (b)(4), alleging that his onetouch verio reflect meter read inaccurately high compared to his feelings and/or normal readings.The complaint was classified based on the customer care agent (cca) documentation.The patient reported that the alleged meter inaccuracy occurred on (b)(6) 2021 at 6:12 pm.The patient reported obtaining an inaccurate high blood glucose reading of ¿122 mg/dl¿ with the subject.The patient manages his diabetes with a fixed dose of insulin (10 units daily after evening meal).The patient denied making any changes to his usual diabetes management regimen as a result of the alleged issue.The patient reported that 15 minutes after the alleged issue occurred , he developed symptoms of ¿eye pain like a migraine, weak and shaking;¿ symptoms he associated with hypoglycemia.The patient claimed he treated himself with sugar in response to the symptoms and felt better a few minutes afterwards.The patient denied using any other device to test his blood glucose.At the time of troubleshooting, the cca confirmed the unit of measure was set correctly on the subject meter.The cca noted the patient did not have control solution available to perform a quality control test.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed symptoms suggestive of a serious injury adverse event after the alleged product issue occurred.The subject meter could not be ruled out as a cause or contributor to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO REFLECT METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key11224176
Report Number3008382007-2021-04106
Device Sequence Number14327812
Product Code NBW
UDI-Device Identifier00353885011709
UDI-Public00353885011709
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K193475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number023-878
Device Lot Number4686906
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2021
Initial Report FDA Received Date01/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
-
-