| Model Number |
0375704500 |
| Medical Device Problem Code |
Insufficient Cooling (1130)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the device became hot during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: i have a shaver hand piece that is defective at surg ctr- acct# 165128 west health surgery center and am hoping we can ship them a replacement.The device began to get hot and make a clicking noise.This was tied to a recent placement agreement that was signed on 10/1, i have attached the document here.Below is the serial number on the defective handpiece.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) excessive force applied to the cutter/burr by the user with poor or no suction during use.(excessive force may cause friction due to bur shaft assembly and housing assembly interaction.) (2) tissue blocking the suction pathway would prevent hot fluid from being removed from the joint.(3) poor arthroscopy pump suction.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the device became hot during use.
|
| |
|