Model Number 0375704500
Device Problems Insufficient Cooling (1130); Overheating of Device (1437)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the device became hot during use.
 
Manufacturer Narrative
Alleged failure: i have a shaver hand piece that is defective at surg ctr- acct# 165128 west health surgery center and am hoping we can ship them a replacement.The device began to get hot and make a clicking noise.This was tied to a recent placement agreement that was signed on 10/1, i have attached the document here.Below is the serial number on the defective handpiece.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be 1) excessive force applied to the cutter/burr by the user with poor or no suction during use.(excessive force may cause friction due to bur shaft assembly and housing assembly interaction.) (2) tissue blocking the suction pathway would prevent hot fluid from being removed from the joint.(3) poor arthroscopy pump suction.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event Description
It was reported that the device became hot during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGING, FORMULA SHAVER (HAND CONTROL)
Type of Device
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11225110
MDR Text Key229612629
Report Number0002936485-2021-00051
Device Sequence Number1
Product Code GFA
UDI-Device Identifier07613327057638
UDI-Public07613327057638
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0375704500
Device Catalogue Number0375704500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Manufacturer Received 12/28/2020
Initial Date FDA Received01/25/2021
Supplement Dates Manufacturer Received12/28/2020
Supplement Dates FDA Received04/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-