The reported event was confirmed however the cause was unknown.Visual evaluation of the returned sample noted one opened (without original packaging), used skylite nitinol stone basket with label.Visual inspection of the sample noted that the green thumb slider had fallen out of the handle.However, the green slider and the handle did not seem damaged.When the basket was reassembled, the basket still could not be opened or closed.This does not meet the specification as "assembly pattern must comply with drawing and visual aid".A potential root cause for this failure could be ¿lack of epoxy /adhesive / cure time¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿insertion 1.Inspect the device prior to use and during the procedure for integrity and function.2.Make sure the basket is closed by retracting (pulling back) the basket tip into the sheath with the thumb slide as shown in figure b.3.With the basket closed, and using the optional introducer provided, carefully advance the distal portion of the closed device through the endoscope until it emerges out of the end of the endoscope.Capture and removal 1.Under direct vision or fluoroscopic guidance, slowly advance the basket tip past the object.2.Open the basket by pushing the thumb slide forward.(refer to figure b).3.Pull the basket backward toward the object while slowly rotating the basket as necessary.4.Once the object has been captured, partially close the basket to secure the object for removal by carefully pulling the thumb slide back.(refer to figure b).5.Slowly remove the basket and stone from the urinary tract.6.If the object is too large, you may need to simultaneously withdraw the basket and the ureteroscope from the urinary system.Directions for disassembly if handle disassembly is desired or required: 1.Squeeze bottom handle half at indicated points and pull down to remove handle bottom.2.Loosen thumbscrew until basket drive wire moves freely.3.Slide sheath and handle assembly over and away from drive wire." the actual/suspected device was inspected.
|