| Model Number |
4FC12 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The 4fc12 sheath with lot number 0010396178 was returned and analyzed.Visual inspection of the sheath showed the shaft was kinked at four locations.It was noted the sheath was likely kinked during shipment.A balloon test catheter under vacuum and with the push button in the forward position failed the insertion into the returned sheath at the kink locations.The afapro28 balloon catheter with lot number 13102 also failed the insertion into the returned sheath at the kink locations.In conclusion, the sheath failed the returned product inspection due to shaft kinks.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The sheath subsequently tested out of specification per the manufacturers investigation.
|
| |
|