• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SUBCUTANEOUS, UNIPOLAR LEAD; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SUBCUTANEOUS, UNIPOLAR LEAD; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 6996SQ41
Medical Device Problem Codes Low impedance (2285); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: qty.2 model: 4968-35, lead; implanted: (b)(6) 2011 model: 4965-35, lead; implanted: (b)(6) 1998 model: dtba1d1, cardiac resynchronization therapy defibrillator (crt-d); implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) defibrillation coil leads and the superior vena cava (svc) defibrillation coil lead all display falling and low impedance.It was noted that the patient was in bad heart failure awaiting transplant and that the impedances were most likely due to fluid between the coils.The defibrillation coil leads remain in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUBCUTANEOUS, UNIPOLAR LEAD
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11226889
Report Number2649622-2021-01580
Device Sequence Number3360676
Product Code LWS
UDI-Device Identifier00613994568014
UDI-Public00613994568014
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/08/2017
Device Model Number6996SQ41
Device Catalogue Number6996SQ41
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2021
Initial Report FDA Received Date01/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6996SQ58, LEAD / 6996SQ58 , LEAD / 672625, ADAPTOR
Patient Age24 YR
-
-