| Model Number |
IPN044066 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the "2411-03 infant high flow cannula was found to be pulling apart near the prongs".No medical intervention required.No patient injury or desaturation reported.User replaced device with a new device.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The sample was sent to the manufacturer (sunmed) for investigation.The manufacturer reports: "the returned unit's facepiece was found to have a separation occurring at the seam.There are multiple variables, both through the manufacturing process, as well as in the end user environment that could contribute to this failure mode.A look at the batch records confirms that this appears to be an isolated incident with lot 012020 and furthermore, the tubing thickness was measured and found to be within specification." based on the investigation performed, the complaint was confirmed; however, a root cause could not be established.
|
| |
|
Event or Problem Description
|
|
It was reported that the "2411-03 infant high flow cannula was found to be pulling apart near the prongs".No medical intervention required.No patient injury or desaturation reported.User replaced device with a new device.Patient condition reported as "fine".
|
| |
|