ARGON MEDICAL DEVICES TRANSJUGULAR LIVER ACCESS BIOPSY SYSTEM; TLAB Back to Search Results
Model Number TL-19
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the images provided of the incident, the complaint can be confirmed.Argon has recalled tlab products for the 7f introducer sheath distal tip separating during procedures.Lot number 1451428 used in this incident was included in the recall that went into effect december 2020.It is unknown at this time why the complainant facility still had the product in its possession and used after the recall notifications were sent to customers.The manufacturing defect was identified before argon started receiving the complaints that prompted the recall.(b)(4) was performed and it adequately addressed the root causes and corrective/preventative actions taken.All affected lots were manufactured before the tip bonding process was updated.Corrective actions included modifying the tipping mandrels to shorten the narrowed tip to 0.300¿, adding band markings at 0.100¿ and 0.200¿ onto the mandrel to simplify the locating of the mandrel within the introducer sheath tube and aid in the alignment of the tube and mandrel within the tipping die by the operator.It also included adding a 100% bend test as well as 100% visual inspection for wrinkling to the inspection process.
 
Event or Problem Description
Radio-opaque band on the tip of the sheath broke off in the patient and traveled to a pulmonary artery in the patient.It's in a periphery artery and was left in the patient.Patient is fine and has no issues on test.
 
Event or Problem Description
Follow up.
 
Additional Manufacturer Narrative
H3 other text: placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSJUGULAR LIVER ACCESS BIOPSY SYSTEM
Common Device Name
TLAB
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX
MDR Report Key11227588
Report Number1625425-2021-00835
Device Sequence Number17163427
Product Code DYB
UDI-Device Identifier00886333218516
UDI-Public00886333218516
Combination Product (Y/N)N
Initial Reporter StateNY
PMA/510(K) Number
K022634
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/05/2023
Device Model NumberTL-19
Device Catalogue NumberTL-19
Device Lot Number1451428
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/25/2021
Supplement Report FDA Received Date02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-