ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported complaint of the autopulse platform (serial #(b)(4)) power off by itself was not confirmed during functional testing.The autopulse platform powers on and doesn't power off by itself.The initial functional testing failed, the autopulse platform displayed user advisory "(ua)41" (patient temperature sensor failure) error message upon powering on, unrelated to the reported complaint.The root cause of ua41 error was due to a defective temperature sensor likely due to defective component.The temperature sensor was replaced to remedy ua41.No physical damaged observed during visual inspection.During archive data review, observed multiple ua41 (patient temperature sensor failure) error messages in the archive, unrelated to the reported complaint.After service repair completion, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The autopulse passed all functional testing.
 
Event or Problem Description
During reanimation training on a manikin, customer reported that the autopulse platform (serial #(b)(4)) power off by itself.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Common Device Name
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
MDR Report Key11228303
Report Number3010617000-2021-00042
Device Sequence Number3927760
Product Code DRM
UDI-Device Identifier00849111000536
UDI-Public00849111000536
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2020
Initial Report FDA Received Date01/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-