• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH PLATE ANCHORAGE MTP / VERSION V2 - RIGHT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH PLATE ANCHORAGE MTP / VERSION V2 - RIGHT; PLATE, FIXATION, BONE Back to Search Results
Model Number PLP11342
Medical Device Problem Codes Material Fragmentation (1261); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device remains in the patient.
 
Event or Problem Description
As reported: "primary procedure, right first mtp.It was reported that when the drill guide was seated into a hole of the plate and after drilling was done, a thread-like piece of metal was seen in the patient.The piece was removed by the surgeon.Surgeon reported that it felt as though the drill guide did not seem to be seating as deeply in the hole as it had in the previous holes, and believes that the thread of metal was shaved off of the plate while drilling.Procedure was completed successfully with no delay.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATE ANCHORAGE MTP / VERSION V2 - RIGHT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11229114
Report Number0008031020-2021-00029
Device Sequence Number3512951
Product Code HRS
UDI-Device Identifier07613252260516
UDI-Public07613252260516
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K083447
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPLP11342
Device Catalogue NumberPLP11342
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2021
Initial Report FDA Received Date01/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-