| Model Number |
115-547 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified underweight, broken shell, crease fold.A microscopic analysis was performed which identify sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on anterior assessed as an unidentified (tear) opening.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: right side implant exchange with no complaint against the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side implant exchange with no complaint against the device.Healthcare professional later noted rupture.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side implant exchange with no complaint against the device.Healthcare professional later noted rupture.Device has been explanted.
|
| |
|